基本誤差
測量氣體
|
測量范圍
|
基本誤差
|
報警誤差及報警設置范圍
|
響應時間
|
CH4
|
(0.00~1.00)%CH4
|
±0.10%CH4
|
報警設置范圍:
0.5~4.00%CH4
報警點:1.00% CH4
報警誤差:±0.05% CH4
|
20s
|
(1.00 ~3.00)%CH4
|
真值的±10%
|
|||
(3.00~4.00)%CH4
|
±0.30%CH4
|
|||
O2
|
(0~30)% O2
|
測量范圍的±3.0%
|
報警設置范圍:
16% O2~19.5% O2
報警點:18% O2
報警誤差:±0.1% O2
|
20s
|
CO
|
(0~100)ppm CO
|
不大于±(1.5+2.0%真值)
|
報警設置范圍:
25~100ppmCO
報警點:24ppmCO
報警誤差:±1ppmCO
|
60s
|
(100~500)ppmCO
|
不大于±4.0%真值
|
|||
>500)ppmCO
|
不大于10.0%真值
|
|||
H2S
|
(0~49)ppm H2S
|
±3ppmH2S
|
報警設置范圍:
5ppm~15ppmH2S
報警點:10ppm
報警誤差:±3ppmH2S
|
45s
|
(50~100)ppm H2S
|
真值的±10%
|